Automatic Contrast Delivery System Specificaion
BASIC SPECIFICATION |
Flow Rate
HEAD A (Contrast Medium) 0.1 ml - 10ml/sec (0.1 ml increments)
HEAD B (Saline) 0.1 ml - 10ml/sec (0.1 ml increments) |
Volume
HEAD A (Contrast Medium) 0.1 ml - 10ml/sec (0.1 ml increments)
HEAD B (Saline) 0.1 ml - 10ml/sec (0.1 ml increments) |
| Preassure Limit |
300 PSI (2058kpa) |
| Area Memory |
255 Maximum |
| Remaining Volume Display |
Bar Graph and a number |
| Scan time |
Yes |
| Multiple Injection |
Yes (Up to 5 Phase Injection) |
| Auto Return Function |
Yes |
| Warning Function |
Yes |
| Interval Time |
Abritary Setup is possible |
| Preassure Display |
Yes (Chart Graph) |
| Small Volume Injection |
Electric Knob (Job Dial) |
| Dispalay Screen |
Color Liquid Crystal |
| Record of Injection Report
| 20 |
| Test Injection
| Yes |
| Power Source
| AC 100V (Japan), AC117V/AC 230V (Overseas) |
| Power Consumption
| 250 VA |
Safety Device
Flow Rate: An Error message is displayed when the flow rate reaches At 125%.
Volume: An Error message is displayed when the injection
duration exceeds the setup injection time by 125%.
Preasure Limit: Safety Device is actuated at the setup value
of +/- 30 P.S.I (+/-207kpa) and a warning message is
dispalyed when the actual preassure value.
Hardware error: An error message is displayed when the
connector etc. is extracted.
|
| Self Checking Function
| Yes |
| Emergency Stop Button
| Powerhead - 3, Console - 1, Hand-Switch - 1 |
| Adaptable Syringe |
100ml and 200ml disposable syringes for contrast medium
100ml disposable syringes for Saline.
|
| |
|
| |
|
| |
|